THE ORCBS > Occupational Safety > Programs & Guidelines > Confined Space > Confined Space Entry Program >

Appendix B. Atmospheric Testing And Monitoring

Procedure For Atmospheric Testing And Monitoring

Atmospheric testing is necessary for two purposes: evaluation of the hazards of the permit space and verification that acceptable entry conditions for entry into that space exist.

  1. Evaluation Testing

    The atmosphere of a confined space should be analyzed using equipment of sufficient sensitivity and specificity to identify and evaluate any hazardous atmospheres that may exist or arise, so that appropriate entry procedures can be developed and acceptable entry conditions stipulated for that space.

    A minimum of three tests should be performed to identify atmospheric hazards in confined spaces. These tests must be performed in the following sequence:

    • Oxygen Content
    • Flammability
    • Toxicity

  2. Verification Testing

    The atmosphere of a permit space which may contain a hazardous atmosphere should be tested for residues of all contaminants identified by evaluation testing using permit specified equipment to determine that residual concentrations at the time of testing and entry are within the range of acceptable entry conditions.

  3. Duration of Testing

    Measurement of values for each atmospheric parameter should be made for at least the minimum response time of the test instrument specified by the manufacturer.

  4. Testing Stratified Atmospheres

    When monitoring for entries involving a descent into atmospheres that may be stratified, the atmospheric envelope should be tested a distance of approximately 4 feet in the direction of travel and to each side. If a sampling probe is use, the entrant's rate of progress should be slowed to accommodate the sampling speed and detector response.

  1. Equipment Calibration

    To ensure that the atmospheric testing equipment is functioning properly, any direct reading test device should not be used without performing the following three operations: 

    • Inspection
    • Calibration
    • Function Test

    All three operations should be performed according to specific manufacturers instructions.

AIR MONITORING GUIDE

  1. Calibrate Instrument
  2. Inspect Instrument
    • Check physical condition of instrument (case, meter, attachments, hoses for cracks)
    • Review instructions to insure you know how to use the device and interpret results.
  3. Perform Function Test
    • Oxygen sensor: breath into sampling device to reduce the oxygen level below 19.5%. The oxygen alarm should sound.
    • Combustible gas sensor: remove cap of solvent magic marker or open a cigarette lighter without a flame near the sampling device until it reaches a 10% reading. The combustible gas sensor should sound.
    • Always perform a function test in the field before use.
    • Never perform a function test in the suspected atmosphere.
  4. Pre-Test Space
    • Zero instrument in known fresh air.
    • Test entire space, top to bottom, every four feet and in the direction of travel.
    • Order of tests:
      1. Oxygen
      2. Flammability
      3. Toxicity
  5. Monitor the Space
    • If continuous monitoring is required, position the instrument near the workers breathing zone.

If any of the alarms sound, exit the space immediately.

Always record your readings.

Contact the ORCBS if any atmospheric hazards cannot be reduced below the PEL with ventilation.

Confined Space Table of Contents

 

Office InformationRadiation SafetyChemical SafetyBiological SafetyOccupational Safety
Animal Handler
Hazardous Waste EnvironmentalTrainingManuals & FormsMSDSLinks

orcbs@msu.edu • MSU Home PageOffice of Vice President for Research and Graduate Studies

© MSU Board of Trustees, All Rights Reserved